Verification of analytical performance of the 25-OH vitamin D assay on the Abbott Alinity ci®: Experience of the central laboratory of Mohammed VI University Hospital of Oujda
1 Mohammed First University, Faculty of Medicine and Pharmacy of Oujda, Morocco.
2 Biochemistry laboratory, Mohammed VI University Hospital of Oujda, Morocco.
Research Article
World Journal of Biology Pharmacy and Health Sciences, 2024, 20(01), 476–482.
Article DOI: 10.30574/wjbphs.2024.20.1.0727
Publication history:
Received on 17 August 2024; revised on 15 October 2024; accepted on 17 October 2024
Abstract:
The primary goal of our study was to verify the analytical performance of the 25-OH Vitamin D measurements conducted on the Abbott CI analyzer using the immuno-chemiluminescence method. This verification took place in the biochemistry laboratory at Mohammed VI University Hospital in Oujda.
Our methodology was aligned with the recommendations outlined in the French accreditation committee (COFRAC) technical guide (GTA) 04, emphasizing the assessment of reproducibility and repeatability.
Overall, the results of this evaluation were satisfactory and met the standards set by both the supplier and the French Society of Clinical Biology. This study illustrates that the biochemistry laboratory at Mohammed VI University Hospital of Oujda is equipped to provide accurate and precise results, which are crucial for effective clinical diagnosis and decision-making.
Keywords:
25-OH Vitamin D assay; Analytical performance; Repeatability; Reproducibility; Alinity CI analyzer; Immuno-chemiluminescence