Verification of analytical performance of Complement C4 on the Abbott Alinity ci®: Experience of the central laboratory of Mohammed VI University Hospital of Oujda
1 Mohammed First University, Faculty of Medicine and Pharmacy of Oujda, Morocco.
2 Biochemistry department, Central laboratory of Mohammed VI University Hospital, Oujda, Morocco.
Research Article
World Journal of Biology Pharmacy and Health Sciences, 2024, 20(03), 623-629.
Article DOI: 10.30574/wjbphs.2024.20.3.1025
Publication history:
Received on 14 November 2024; revised on 22 December 2024; accepted on 25 December 2024
Abstract:
The primary objective of our study was to validate the analytical performance of complement C4 assay performed on the Abbott Alinity ci®: analyzer using the Immunoturbidimetric method. This validation was conducted in the biochemistry laboratory at Mohammed VI University Hospital in Oujda. The methodology followed the guidelines outlined in the french accreditation committee (COFRAC) technical guide (GTA) 04, focusing on the evaluation of reproducibility and repeatability.
Overall, the results were satisfactory and met the standards set by both the manufacturer and the French Society of Clinical Biology. This study highlights the capability of the biochemistry laboratory at Mohammed VI University Hospital in Oujda to deliver accurate and reliable results, which are critical for effective clinical diagnosis and decision-making.
Keywords:
Complement C4; Verification; Repeatability; Reproducibility; Alinity CI
Full text article in PDF:
Copyright information:
Copyright © 2024 Author(s) retain the copyright of this article. This article is published under the terms of the Creative Commons Attribution Liscense 4.0