Verification of the analytical performance of the serum magnesium assay on Abbott Architect ci8200

Oumaima Nassiri 1, 2, *, Sabrina Belmahi 1, 2, Abdessamad Amrani 1, 2, El-houcine SEBBAR 1, 2 and Mohammed Choukri 1, 2

1 Faculty of Medicine and Pharmacy of Oujda, Mohammed First University, Morocco.
2 Central Laboratory, Mohammed VI University Hospital of Oujda, Morocco.
 
Research Article
World Journal of Biology Pharmacy and Health Sciences, 2024, 18(02), 359–364.
Article DOI: 10.30574/wjbphs.2024.18.2.0282
 
Publication history: 
Received on 02 April 2024; revised on 12 May 2024; accepted on 14 May 2024
 
Abstract: 
Magnesium is recognized as the fourth most important cation in the body and the second most abundant intracellular cation, following potassium. Its physiological significance lies in its capacity to form complexes with crucial intracellular anionic ligands, notably ATP, and its ability to compete with calcium for specific binding sites on proteins and cell membranes. Over 300 enzyme reactions rely on magnesium. The concentration of free magnesium within cells determines the magnesium content of these enzymes. Magnesium influences myocardial contraction and the electrical activity of myocardial cells, as well as the specialized conduction systems of the fetus, by modulating the transmembrane movements of ions like sodium, potassium, and calcium. Furthermore, magnesium can affect the contractility of vascular smooth muscle, and alterations in intracellular magnesium levels may impact cell proliferation or contraction. Magnesium is involved in nucleic acid and protein metabolism, intermediary metabolism, energy production, and consumption reactions, and it plays specific roles in various physiological systems such as the neuromuscular and cardiovascular systems.
Our research aims to assess the analytical accuracy of the serum magnesium assay method utilizing an Abbott kit on the Architect ci8200 automated system in the biochemistry laboratory of Mohammed VI University Hospital of Oujda. The evaluation of the kit's analytical accuracy was conducted within the flexible scope A by conducting a performance analysis on the Architect ci8200. We examined repeatability, reproducibility, measurement imprecision, and compared results between two Architect ci8200 instruments. All findings align with the acceptability criteria recommended by the supplier and the Valtec protocol of the Société Française de Biologie Clinique (SFBC), indicating overall satisfactory outcomes for the study.
The Architect ci8200 automated system demonstrated satisfactory analytical performance for reliably determining serum magnesium levels. Method verification in medical laboratories is essential to guarantee the accuracy, precision, and reliability of test results. Verification entails confirming that the test method used is suitable for its intended purpose, yields result consistent with the claimed performance characteristics, and meets the laboratory's quality control and quality assurance standards.
 
Keywords: 
Magnesium; Verification; Quality; Architect ci 8200; Repeatability; Reproducibility
 
Full text article in PDF: